Group, shower screen, portafilter and leaks

12 questions

What is the correct water flow rate through the groups?

As a reference, the group flow rate without a portafilter fitted should be approximately 100 cm³ in 8–10 seconds.

Visually, when activating the group without a portafilter, a uniform flow should be observed across the entire shower screen. This may appear as a rainfall effect or may merge into a central stream. The important point is that the water is distributed evenly.

If thin isolated streams, uneven jets, irregular dripping or dry areas are observed, there may be an issue within the group. When the water is not distributed evenly across the entire coffee puck, the extraction is affected, which has a direct impact on the cup quality.

The most common causes are a dirty or blocked shower screen, a partially obstructed diffuser, a dirty or worn jet, the build-up of limescale or coffee oils, a deformed shower screen or incorrect assembly.

In these cases, it is advisable to replace the shower screen and, consequently, also replace the group gasket/seal.

To maintain a stable and consistent flow rate through the group, it is essential to follow a proper group cleaning routine.

Water keeps flowing into the drain tray. What could be the cause?

If the machine is continuously discharging water into the drain tray, the first thing to bear in mind is that several circuits may discharge into this tray. It does not only receive water from the groups: water may also come from the expansion-retention valve and, on some models, from the boiler drain valves. Therefore, before diagnosing the fault, it is necessary to identify which circuit the water is actually coming from.

1. Water appears only during coffee extraction

If water is discharged into the drain tray while coffee is being made, a common cause is that the group 3-way solenoid valve is not closing correctly and part of the water is being diverted to the drain instead of flowing through the coffee.

2. Water drips while the machine is idle

If the machine is not dispensing coffee and water is still dripping into the drain tray, it may be due to:

  • a group discharge valve that is not closing properly,
  • dirt or limescale inside the group solenoid valve,
  • or an internal leak that eventually drains into the waste outlet.

3. Water appears during filling or while the pump is running

In this case, the expansion-retention valve should also be considered.

A small amount of dripping from the expansion valve may be normal while it regulates pressure, but if the discharge is excessive or continuous, it is advisable to check:

  • the pump pressure,
  • the expansion valve setting
  • and the overall behaviour of the hydraulic circuit.

4. The machine has a boiler drain connected to the drain tray

On some models, the boiler drain valves also discharge into this area. If there is leakage or improper sealing at this point, water may appear in the drain tray even though the fault is neither in the group nor in the expansion valve.

5. What should be checked first?

Before dismantling any components, it is advisable to check:

  • whether the water appears only during coffee extraction or also when the machine is idle,
  • whether it coincides with the machine filling cycle,
  • whether it affects one group or all groups,
  • and whether the discharge appears to come from the group, the expansion valve or a drain line.

6. What does it mean if it never stops?

If water flows continuously, the most likely causes are:

  • a group solenoid valve that is not closing correctly,
  • an expansion-retention valve discharging excessively,
  • or a leak/improper sealing in a boiler drain circuit.

Suitable gasket types if the original one does not fit properly

If the original group gasket does not fit properly or fails to provide a correct seal, the issue may not be with the gasket itself, but rather with wear on the portafilter lugs or on the group head locking surfaces. In such cases, it may be necessary to fit a thicker gasket or add an additional spacer gasket to increase the overall thickness and restore a proper seal. In addition to sealing performance, it is important to check that the portafilter is not over-rotated. The handle axis should not extend beyond the perpendicular position relative to the front of the machine in order to maintain proper ergonomics. If the problem persists after fitting a thicker gasket, replacement of the portafilter is recommended.

No water is coming out of one group, but the others are working. What should be checked?

If no water is coming out of one group while the others are operating normally, the problem is usually located within the affected group itself rather than in the machine’s general water supply system.

First, check whether the group is correctly receiving the command from the keypad. Then inspect the shower screen, diffuser, group solenoid valve, hydraulic circuit, group jet (chicleur) and the flowmeter associated with that group.

If the fault moves to the opposite side after swapping keypads or electrical connections, the problem is located in the keypad or the wiring.

If the fault remains on the same group, troubleshooting should focus on that specific group and its associated electrical and hydraulic circuits.

How to clean portafilters?

Associated symptoms

  • Rancid or consistently bitter coffee taste.
  • Clogged filters or irregular flow.
  • Coffee grounds in the cup.
  • Difficulty removing built-up residue.

Possible causes

  • Buildup of coffee oils and residues.
  • Cleaning only with hot water.
  • Extended use without chemical cleaning.
  • Deformed or worn filter baskets.

Checks / observations

  • Remove the filter basket from the portafilter.
  • Soak only the metal parts in hot water with specific detergent.
  • Do not immerse handles if made of plastic or wood.
  • Allow detergent to act for the recommended time.
  • Rinse thoroughly with clean water.
  • Dry before reassembling.
  • Ensure filter holes are clear and uniform.
  • Replace deformed or heavily worn filters.

How should groups be cleaned?

Associated symptoms (when not properly cleaned)

  • Bitter or burnt coffee taste.
  • Unpleasant odors during extraction.
  • Irregular water flow from the shower screen.
  • Dripping from the group when idle.
  • Visible buildup of coffee residues.

Possible causes

  • Lack of daily group cleaning.
  • No detergent backflushing cycles performed.
  • Cleaning with water only.
  • Shower screens and valves clogged with coffee residues.

Checks / observations

  • Daily cleaning (operator):
    • Insert blind filter into portafilter.
    • Run group cleaning cycles.
    • Repeat with water only for rinsing.
    • Clean the external shower screen after service.
  • Periodic cleaning (deep):
    • Remove shower screen and gasket.
    • Clean with brush and specific detergent.
    • Rinse thoroughly before reassembly.
    • Replace gasket and shower when removed.

Proper cleaning prevents most group failures and significantly improves cup quality.

The shower screen and gasket have fallen off. What should be checked?

Associated symptoms

  • The shower screen and/or gasket fall out when removing the portafilter.
  • The portafilter does not fit properly after reassembly.
  • Water leaks around the portafilter edge.
  • Irregular coffee extraction or splashing.

Possible causes

  • Hardened, deformed or excessively worn group gasket.
  • Shower screen incorrectly positioned or not centered.
  • Buildup of coffee residues between the shower screen and group.
  • Use of incompatible shower screens or gaskets.
  • Incorrect reassembly after cleaning.

Checks / observations

  • Check gasket condition: if hard, cracked or deformed, replace it.
  • Clean the group seat thoroughly before reassembly.
  • Install the shower screen centered, then the gasket (depending on group design).
  • Ensure the shower screen sits completely flat.
  • Insert the empty portafilter and verify it locks firmly without play.
  • If the shower screen falls again, always replace the gasket.

Steam/water lever is stiff, does not return or does not shut off completely. What could be the cause?

Symptom: lever feels stiff, does not return smoothly or does not fully shut off.

Cause: lack of food-grade lubrication on cam and shafts, worn seals in the mechanism, or limescale/dirt buildup in the assembly.

Observations: do not force the mechanism; disassemble the assembly, clean it, replace seals and lubricate with food-grade grease. If dripping is also present, it is often related to worn seals or valve seat.

Group shower screen has irregular flow or sprays. What should be done?

Symptom: uneven water pattern, with jets and dry spots.

Cause: shower screen or diffuser clogged with coffee residues and oils; partially blocked diffuser or deformed shower screen.

Observations: remove the shower screen, clean it (brush + specific cleaner), clean the diffuser and internal channels, and perform a flush. If the shower screen is marked or warped, replace it.

Water comes out from the back of the group or drains during extraction. What does it indicate?

Symptom: during extraction, water flows to the drain/return or an internal leak appears behind the front panel.

Cause: discharge remains open: faulty 3-way solenoid valve or valve seat/seal not sealing properly; may also be caused by overpressure due to blockage in the shower screen or coffee puck.

Observations: check if pump pressure behaves abnormally and coffee barely flows; clean shower screen and diffuser, test without coffee (blind portafilter/diagnostic). If water flows to the drain during extraction, it indicates the 3-way solenoid valve is not functioning correctly.

The group drips when not extracting. What are the typical causes?

Symptom: continuous dripping from the shower screen while the machine is idle.

Cause: 3-way solenoid valve (discharge) not closing properly (due to dirt/limescale) or mechanical group valve with a worn seat.

Observations: often increases after descaling or with hard water; perform cleaning/flush; if not resolved, disassemble and clean or replace the valve core, stem or complete solenoid valve assembly.

Water leaks from the portafilter during extraction. What is the usual cause?

Symptom: leakage around the portafilter rim during extraction.

Cause: worn or hardened group gasket, or improperly seated portafilter (worn lugs).

Observations: check and clean the group seat and shower screen; test with another portafilter or basket; if the issue persists, replace gasket and shower screen. If it only leaks with one portafilter, the issue is with that portafilter or its basket.